MGH/MF Waveform Database 1.0.0

File: <base>/mgh014.hea (1,656 bytes)
mgh014 8 360/0.476 1782000 19:19:24 07/11/1991
mgh014.dat 212 364(-346)/mV 12 0 -307 4164 0 ECG Lead I
mgh014.dat 212 348(-136)/mV 12 0 -119 47267 0 ECG Lead II
mgh014.dat 212 324(-113)/mV 12 0 -93 34161 0 ECG Lead V
mgh014.dat 212 19.7(-2454)/mmHg 12 0 2047 28053 0 ART
mgh014.dat 212 1000 12 0 1975 55007 0 OFF
mgh014.dat 212 1000 12 0 2047 36134 0 OFF
mgh014.dat 212 1000 12 0 -404 23060 0 Resp Inp.
mgh014.dat 212 1000 12 0 -1911 39567 0 CO2
#<age>: 58 <sex>: M <diagnoses>: Coronary artery bypass graft
# PHARMACOLOGIC SUPPORT:
#   Dopamine @ 150 mcg/min
# GENERAL COMMENTS:
#   Stop/FFW @ 70 min - 71 min
# ELECTROCARDIOGRAPHIC DATA:
# UNDERLYING RHYTHM:
#   Atrial pacing @ 95 bpm
# RHYTHM DISTURBANCES:
#   Unifocal ventricular ectopy
# PACEMAKER DATA:
#   Temporary atrial fixed rate pacing (AOO)
#   Epicardial wires
#   Medtronic 5330 pulse generator
#   MA 15
# ECG INTERPRETATION:
#   Anterior infarct
#   ? subendocardial apico-lateral infarct
#   Nonspecific ST segment and T wave abnormalities consistent with digitalis effect or ischemia
# TECHNICAL COMMENTS:
#   Chest lead = V3
# HEMODYNAMIC DATA:
#   ART: 114/58   	MEAN: 84
#   PAP: 41/13
#   RAP: 13
# INTRA-AORTIC BALLOON DATA:
#   40 cc balloon
#   Aries console
#   1:8 balloon ratio (1: 1.1 :2.1:4 during timing sequence)
# WAVEFORM INTERPRETATION:
#   Radial and aortic lumen IABP traces are recorded.
#   Timing sequences @ 14 min- 18 min and 29 min-33 min
#   Late deflation and early inflation during timing sequence
# TECHNICAL COMMENTS:
#   Zero to RAD lumen @ 11 min
#   Zero to AO lumen @ 12 min
# RESPIRATORY DATA:
# RATE:   18 bpm
# MODE OF VENTILATION:
#   Spontaneous